Test 3: Channels 1 and 2 Frequency Sensitivity (Agilent 53131A/132A
Only)
1-26 Assembly-Level Service Guide
1
200 MHz to 225 MHz Sensitivity for Agilent 53131A/132A
1 Leave the signal generator connected to Channel 1. Change the signal generator
settings to 225 MHz at 30 dBm.
2 On the Counter, press Run key.
3 Increase the power level by 0.1 dB steps until the Counter displays a stable
reading of 225 MHz.
The Counter should read 225 MHz when the input signal level is
15.0 dBm (40 mVrms).
4 Sweep the signal generator frequency from 225 MHz to 200 MHz.
The Counter should read frequencies from 225 MHz to 200 MHz when the input
signal level is 15.0 dBm (40 mVrms).
5 Record the actual reading in the Performance Test Record (Test 3, Line 5).
6 Connect the signal generator to Channel 2.
7 On the Counter, press Freq & Ratio key until Frequency 2 is displayed.
8 Repeat steps 1 through 4 for Channel 2.
9 Record the actual reading in the Performance Test Record (Test 3, Line 6).
10 Disconnect the test setup.