Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (Agilent 53131A/132A
Only)
Assembly-Level Service Guide 1-25
1
If you are testing a Counter with front-panel terminals (Standard), record the
reading on Line 2 of the Standard portion of the test record.
If you are testing a Counter with rear terminals (Option 060), record the reading on
Line 2 of the Option 060 portion of the test record.
NOTE The remaining sensitivity procedures are NOT required for Counters that contain
Option 060 Rear Terminals; thus, this completes the sensitivity portion of the
Frequency Performance Test. Go to the Frequency Accuracy Test on page 1-27.
100 MHz to 200 MHz Sensitivity for Agilent 53131A/132A
1 Leave the signal generator connected to Channel 2. Change the signal generator
settings to 200 MHz at 30 dBm.
2 On the Counter, press Run key.
3 Increase the power level by 0.1 dB steps until the Counter displays a stable 200
MHz reading.
The Counter should display 200 MHz when the input signal level is
≤−17.5 dBm (30 mVrms).
4 Sweep the signal generator frequency from 200 MHz to 100 MHz.
The Counter should read frequencies from 200 MHz to 100 MHz when the input
signal level is 17.5 dBm (30 mVrms).
5 Record the actual reading in the Performance Test Record (Test 3, Line 3).
6 Connect the signal generator to Channel 1.
7 On the Counter, press Freq & Ratio key until Frequency 1 is displayed.
8 Repeat steps 1 through 4 for Channel 1.
9 Record the actual reading in the Performance Test Record (Test 3, Line 4).