Instruction manual
Chapter 1 Performance Tests
Agilent 53131A/132A/181A Operational Verification
1-8 Assembly-Level Service Guide
1
Run Self Test
1 Disconnect the input signals from the Counter.
2 Power down (turn off the Counter), press and hold Recall (Utility) key, then press
POWER key.
3 Press Recall key until TEST: ALL ? is displayed.
4 Press Enter key.
Observe that ALL: PASS is displayed after the self test is completed.
If ALL: FAIL is displayed or a failure is indicated, refer to the troubleshooting
section in Chapter 2, “Service,” of this guide.
5 For Agilent 53131A/132A, mark Pass or Fail in the Agilent 53131A/132A
Performance Test Record on page 1-41, Test 4.
5’ For Agilent 53181A, mark Pass or Fail in the Agilent 53181A Performance Test
Record on page 1-71, Test 3.
Termination Check
1Cycle the POWER key to preset the Counter.
2 Connect a digital voltmeter (DVM) to CHANNEL 1 as shown in Figure 1-2, and
set it to measure ohms (set DVM to appropriate Ωrange).
Agilent 53131A/132A is shown in above figure. This setup for the Agilent 53181A requires a connection to
Channel 1 only.
Figure 1-2. Termination Check Setup
Counter
HP 3458A
Digital Multimeter
BNC - to - Dual
Banana Plug