Instruction manual
Chapter 1 Performance Tests
Introduction
1-2 Assembly-Level Service Guide
1
Introduction
This chapter provides procedures to test the electrical performance of the Agilent
53131A/132A Universal Counter and Agilent 53181A Frequency Counter, using the
specifications listed in Chapter 7, “Agilent 53131A/132A Specifications,” and
Chapter 8, “Agilent 53181A Specifications,” of this guide.
Three types of testing are provided:
• Operational Verification
• Complete Performance Tests
• Optional GPIB Verification
This chapter is organized as follows:
Operational Verification
The operational verification is an abbreviated series of checks that may be performed
to give a high degree of confidence that the instrument is operating properly without
performing the complete performance tests. An operational verification is useful for
incoming inspection, routine maintenance, and after instrument repair.
• Introduction page 1-2
• Equipment Required page 1-4
• Agilent 53131A/132A/181A Operational Verification page 1-5
• Agilent 53131A/132A Complete Performance Tests page 1-16
• Agilent 53131A/132A GPIB Verification Program (Optional) page 1-40
• Agilent 53131A/132A Performance Test Record (Page 1 of 6) page 1-41
• Agilent 53181A Complete Performance Tests page 1-47
• Agilent 53181A GPIB Verification Program (Optional) page 1-70
• Agilent 53181A Performance Test Record (Page 1 of 4) page 1-71