Chapter 2 Service
The Agilent 53131A/132A Calibration Procedures
Assembly-Level Service Guide 2-15
2
The Agilent 53131A/132A Calibration Procedures
First Determine the Counter Firmware Revision
Press and hold Recall (Utility) key, then cycle POWER key.
REV: n is displayed; where “n” represents the numeric firmware revision code.
If 3317, 3335, or 3402 is displayed as the firmware revision code, then perform the
calibration procedures in Chapter 6, “Backdating.”
If the Counter’s firmware revision code is higher than the ones in the previous
paragraph, then perform the following procedures in this section.
NOTE After completing the instrument calibration, BE SURE to secure the calibration by
performing the procedure in the section titled “To Secure Against Calibration” on
page 2-13 of this chapter.
To Calibrate the Offset for Channels 1 and 2
1 To access the Calibration menu, power down, and press and hold the Scale &
Offset key, then press POWER key.
The first item in the Calibration menu is a message indicating the calibration security
status: CAL SECURE or CAL UNSECURE.
The Counter calibration must be unsecured to perform calibration.
If CAL UNSECURE is displayed, the Counter is already unsecured for calibration;
go to step 5. However, if CAL SECURE is displayed then perform steps 2 through 4
to unsecure calibration.
2 Press Scale & Offset key until CODE: 0 is displayed.