About the Agilent 53131A/132A Calibration Menu
2-12 Assembly-Level Service Guide
2
Equipment and Counter Setup: See Figure 2-1.
Figure 2-1. Calibration Setup for CAL: TI FINE?
Out of Agilent 8130A, into Agilent 59992A J06 Calibrator Signal:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low: −0.50 V
Input Mode: Normal
Timebase: Any external timebase you provide is ignored during calibration.
Procedure: From the front-panel calibration menu, one keypress invokes the
calibration. You are prompted to press buttons on the Agilent 59992A J06
Calibrator for each part of a four-part calibration.
Note that CAL: TI FINE? requires the completion of four steps in order.