Instruction manual

Chapter 2 Service
About the Agilent 53131A/132A Calibration Menu
2-8 Assembly-Level Service Guide
2
The Agilent 53131A/132A Calibration Menu Tree
NOTE Turn power off, press and hold Scale & Offset key, then press POWER key to access
this menu. This menu does not exist in early versions (firmware revisions 3317,
3335, and 3402) of the Agilent 53131A/132A. In the early versions of the Counter,
the CAL: menu item resides in the Utility menu, and there is no calibration
security capability. See Chapter 6, “Backdating,” in this guide for correct
calibration procedures.
1
This menu item appears and calibration is permitted only if calibration is unsecure. Enter in the correct code
to change calibration to secure.
2
Timebase can be automatically calibrated only if the timebase option is installed.
Scale &
Offset
On / Stby
POWER
CAL SECURE CAL UNSECURE
1
CAL: OFFS2?
CAL: OFFS1?
CAL: GAIN2?
CAL: GAIN1?
CAL: TIMEBAS?
CAL: TI FINE?
CAL: TI QUIK?
CODE: 0
CAL COUNT?
HELP: CAL?
HELP: SECURE?
HELP: CODE?
2