94 Operation and Maintenance Manual
15
Carefully bend the H
2
line into position so that the 1/16" Valco nut and
ferrule can be screwed into the side port of the splitter fitting. Tighten the
connection of the H
2
line to the splitter fitting using a 3/8" wrench on the
vertical flat of the splitter fitting and 1/4" wrench on the Valco nut.
16 Make sure that no other connections have loosened or moved out of
alignment, if so, reposition or retighten the fittings as needed.
17 The assembled Burner is now ready for re-installation on the GC.