Operation and Maintenance Manual 91
7
Position the upper ceramic tube into the long axis of the splitter fitting so
that about 4 mm of it extends past the top of the fitting. Slide the 0.054" ID
double tapered ferrule over the upper ceramic tube. Gently holding these
parts so that neither the ferrule nor the upper ceramic tube slips out of
position, finger-tighten the union fitting onto the splitter fitting.
Figure 26 Proper Ferrule Orientation to the Large Ceramic Tube
Figure 27 Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly