84 Operation and Maintenance Manual
Changing the Oil Mist Filter (RV5)
The oil mist filter on the RV5 pump has two components: the charcoal odor
filter and the oil coalescing filter element. To replace the filters, disassemble
the oil mist filter assembly with the 4 mm long-handled allen wrench
(provided). The smaller charcoal odor filter sits on top of the larger oil
coalescing filter element. It is recommended to replace the oil coalescing filter
element after 90 days of continuous use, however replacement of the charcoal
odor filter is optional. After replacing the filter, re-assemble the filter
assembly and attach it to the pump flange.