Technical data
Table Of Contents
- Contents
- Introduction
- System Description
- Installation
- Overview
- Step 1: Selecting a Location
- Step 2: Unpack and Inspect the Instrument
- Step 3: Set Up the Vacuum Pump
- Step 4: Connect the Power Cord
- Step 5: Install the Dry Compressed Air or O2 Supply
- Step 6: Install the Signal Output Cables
- Step 7: Install the Dual Plasma Burner
- Step 8: Install the Dual Plasma Controller
- Step 9: Install Column Connections
- Step 10: Install the Transfer Line
- Front Panel Controls and Initial Startup
- Operation
- Maintenance
- Pump Maintenance
- Cleaning the Detector
- Changing the Oil Mist Filter (RV5)
- Reaction Cell Cleaning
- Flow Sensor Calibration
- Detector Sensitivity
- Assembling the Dual Plasma Burner for Component Replacement with the SCD
- Assembling the Dual Plasma Burner for Component Replacement with the NCD
- Tube Replacement for the SCD
- Tube Replacement for the NCD
- Troubleshooting
- Index

66 Operation and Maintenance Manual
Dual Plasma Controller Controls
The Agilent Dual Plasma Controller provides easy access to basic settings.
The baseline signal from the SCD and NCD can be
adjusted from 0 to ±1% of the full scale recorder
output using the offset control. (e.g., +10 mV to
-10 mV for a 1 V full scale setting). This control
can be used to offset the background signal.
Figure 22 Dual Plasma Controller Front Panel
Oxidizer and
Hydrogen Control
Knobs
Allow you to adjust the oxidizer and hydrogen flow
rate.
Temperature Control
Knob
Allows you to adjust the Burner temperature.
Selector Control Knob Changes the display to show the current value or
set point for each setting (temperature in °C,
pressure in Torr, oxidizer in sccm, or hydrogen in
sccm).