Technical data
Table Of Contents
- Contents
- Introduction
- System Description
- Installation
- Overview
- Step 1: Selecting a Location
- Step 2: Unpack and Inspect the Instrument
- Step 3: Set Up the Vacuum Pump
- Step 4: Connect the Power Cord
- Step 5: Install the Dry Compressed Air or O2 Supply
- Step 6: Install the Signal Output Cables
- Step 7: Install the Dual Plasma Burner
- Step 8: Install the Dual Plasma Controller
- Step 9: Install Column Connections
- Step 10: Install the Transfer Line
- Front Panel Controls and Initial Startup
- Operation
- Maintenance
- Pump Maintenance
- Cleaning the Detector
- Changing the Oil Mist Filter (RV5)
- Reaction Cell Cleaning
- Flow Sensor Calibration
- Detector Sensitivity
- Assembling the Dual Plasma Burner for Component Replacement with the SCD
- Assembling the Dual Plasma Burner for Component Replacement with the NCD
- Tube Replacement for the SCD
- Tube Replacement for the NCD
- Troubleshooting
- Index

64 Operation and Maintenance Manual
Detector Front Panel Controls
As illustrated in Figure 21, the front panel is divided into three sections: signal
control, display output control, and power control. Each section is described
below. All the front panel LEDs are red in the ON mode and darkened in the
OFF mode except for POWER. This LED toggles green for ON and red for
STANDBY. In STANDBY mode, the front panel display and high voltage to the
photomultiplier tube (PMT) are turned off. The red LED in the STANDBY mode
serves as a reminder that power is still supplied to the instrument.
Power Controls
Figure 21 Front Panel Controls
ON: Turns on front panel displays and high
voltage to the PMT.
STANDBY: Darkens panel and turns off high
voltage to the PMT. Does not shut off main
power.
Turns pump ON (red) and OFF (dark).
Functions in both ON and STANDBY modes.