Technical data
Table Of Contents
- Contents
- Introduction
- System Description
- Installation
- Overview
- Step 1: Selecting a Location
- Step 2: Unpack and Inspect the Instrument
- Step 3: Set Up the Vacuum Pump
- Step 4: Connect the Power Cord
- Step 5: Install the Dry Compressed Air or O2 Supply
- Step 6: Install the Signal Output Cables
- Step 7: Install the Dual Plasma Burner
- Step 8: Install the Dual Plasma Controller
- Step 9: Install Column Connections
- Step 10: Install the Transfer Line
- Front Panel Controls and Initial Startup
- Operation
- Maintenance
- Pump Maintenance
- Cleaning the Detector
- Changing the Oil Mist Filter (RV5)
- Reaction Cell Cleaning
- Flow Sensor Calibration
- Detector Sensitivity
- Assembling the Dual Plasma Burner for Component Replacement with the SCD
- Assembling the Dual Plasma Burner for Component Replacement with the NCD
- Tube Replacement for the SCD
- Tube Replacement for the NCD
- Troubleshooting
- Index
58 Operation and Maintenance Manual
Step 6: Install the Signal Output Cables
Signal output cables are available from Agilent as standard equipment and can
be used with most data systems. Confirm that the output cable supplied is
correct for your system. A standard cable fitted with two crimp lug connectors
is supplied for use with most integrators, recorders or data systems. Attach
the BNC connector end of the recorder cable to the matching output
connector, labeled RECORDER OUTPUT, on the back of the SCD (see
Figure 17).
Standard Cable Connection
The standard recorder cable is connected to the integrator by installing the
red crimp-lug connector to the signal terminal (+) and the black crimp-lug
connector to the negative terminal (-). No additional ground connection is
required.
HP 3390 Series Integrator Cable Connection
The keyed-edge connector on the HP 3390 series cable is attached to the signal
input of HP 3390 series integrators. Note that the connector can only be
installed one way.
HP 3396 Integrator Cable Connection
The jack plug on the HP 3396 cable is attached to the analog signal input
connector at the rear of the integrator. This cable also works with the Agilent
35900 controller.
HP 5890 GC Analog Input Board
A cable made specifically for this input board is available. This board is used
when the SCD signal is input to ChemStation.
Agilent 6890 GC Analog Input Board
A cable made specifically for this input board is available. This board is used
when the SCD signal is input to ChemStation.