112 Operation and Maintenance Manual
Temperature Reading Problems
Normally, the controller displays the burner temperature set-point. For
diagnostic purposes, a switch in the controller (JP2) can be positioned to allow
the display of the actual thermocouple temperature reading. When the actual
temperature and set-point temperature agree, the heater indicator light on the
front of the controller is illuminated. Note that there is a slight offset between
these readings.