Chromatographic Checkout 4
Advanced User Guide 91
To Check FPD Performance (Sample 5188-5953)
To check FPD performance, first check the phosphorus
performance, then the sulfur performance.
Preparation
1 Gather the following:
Evaluation column, HP- 5 30 m × 0.32 mm × 0.25 µm
(19091J- 413)
FPD performance evaluation (checkout) sample
(5188- 5953)
Phosphorus filter
Sulfur filter and filter spacer
4- mL solvent and waste bottles or equivalent for
autoinjector.
2- mL sample vials or equivalent for sample.
Chromatographic- grade isooctane for syringe wash
solvent.
2 Verify the following:
Capillary column adapter installed. If not, install it.
Chromatographic- grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
Empty waste vials loaded in sample turret.
4- mL vial with diffusion cap filled with isooctane and
inserted in Solvent A injector position.
3 Verify that the Lit Offset is set appropriately. Typically, it
should be about 2.0 pA for the checkout method.
4 Install the evaluation column. (See the procedure for the SS,
PP, COC, MMI, or PTV in the Maintenance manual.)
Set the oven, inlet, and detector to 250 °C and bake out for
at least 15 minutes. (See the procedure for the SS, PP,
COC, MMI, or PTV in the Maintenance manual.)
Be sure to configure the column.
Phosphorus performance
5 If it is not already installed, install the phosphorus filter.
6 Create or load a method with the parameter values listed in
Table 13.