User guide

Chromatographic Checkout 4
Advanced User Guide 81
To Check NPD Performance
1 Gather the following:
Evaluation column, HP- 5 30 m × 0.32 mm × 0.25 µm
(19091J- 413)
NPD performance evaluation (checkout) sample
(18789- 60060)
4- mL solvent and waste bottles or equivalent for
autoinjector.
Chromatographic- grade isooctane
2- mL sample vials or equivalent for sample.
2 Verify the following:
Capillary column jet installed. If not, select and install a
capillary column jet.
Capillary column adapter installed. If not, install it.
Chromatographic- grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
Empty waste vials loaded in sample turret.
4- mL vial with diffusion cap filled with isooctane and
inserted in Solvent A injector position.
3 If present, remove any protective caps from the inlet
manifold vents.
4 Install the evaluation column. (See the procedure for the SS,
PP, COC, MMI, or PTV in the Maintenance manual.)
Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS, PP, COC, MMI, or
PTV in the Maintenance manual.)
Be sure to configure the column
5 Create or load a method with the parameter values listed in
Table 11.
Table 11 NPD Checkout Conditions
Column and sample
Type HP-5, 30 m × 0.32 mm × 0.25 µm
(19091J-413)
Sample NPD checkout 18789-60060
Column mode Constant flow