User guide

76 Advanced User Guide
4 Chromatographic Checkout
To Check TCD Performance
1 Gather the following:
Evaluation column, HP- 5 30 m × 0.32 mm × 0.25 µm
(19091J- 413)
FID/TCD performance evaluation (checkout) sample
(18710- 60170)
4- mL solvent and waste bottles or equivalent for
autoinjector
Chromatographic- grade hexane
2- mL sample vials or equivalent for sample
Chromatographic- grade helium as carrier, makeup, and
reference gas
2 Verify the following:
Chromatographic- grade gases plumbed and configured:
helium as carrier gas and reference gas.
Empty waste vials loaded in sample turret.
4- mL solvent vial with diffusion cap filled with hexane
and inserted in Solvent A injector position.
3 Install the evaluation column. (See the procedure for the SS,
PP, COC, MMI, or PTV in the Maintenance manual.)
Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS, PP, COC, MMI, or
PTV in the Maintenance manual.)
Configure the column
4 Create or load a method with the parameter values listed in
Table 10.
Table 10 TCD Checkout Conditions
Column and sample
Type HP-5, 30 m × 0.32 mm × 0.25 µm
(19091J-413)
Sample FID/TCD checkout 18710-60170
Column flow 6.5 mL/min
Column mode Constant flow
Split/splitless inlet
Temperature 250 °C