User guide

Chromatographic Checkout 4
Advanced User Guide 71
To Check FID Performance
1 Gather the following:
Evaluation column, HP- 5 30 m × 0.32 mm × 0.25 µm
(19091J- 413)
FID performance evaluation (checkout) sample
(5188- 5372)
Chromatographic- grade isooctane
4- mL solvent and waste bottles or equivalent for
autoinjector
2- mL sample vials or equivalent for sample
2 Verify the following:
Capillary column jet installed. If not, select and install a
capillary column jet.
Capillary column adapter installed (adaptable FID only).
If not, install it.
Chromatographic- grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
Empty waste vials loaded in sample turret.
4- mL solvent vial with diffusion cap filled with isooctane
and inserted in Solvent A injector position.
3 Install the evaluation column. (See the procedure for the SS,
PP, COC, MMI, or PTV in the Maintenance manual.)
Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS, PP, COC, MMI, or
PTV in the Maintenance manual.)
Be sure to configure the column.
4 Check the FID baseline output. The output should be
between 5 pA and 20 pA and relatively stable. (If using a gas
generator or ultra pure gas, the signal may stabilize below
5 pA.) If the output is outside this range or unstable, resolve
this problem before continuing.
5 If the output is too low:
Check that the electrometer is on.
Check that the flame is lit (“To light the FID flame” on
page 300).
6 Create or load a method with the parameter values listed in
Table 9.