User guide
Chromatographic Checkout 4
Advanced User Guide 69
To Prepare for Chromatographic Checkout
Because of the differences in chromatographic performance
associated with different consumables, Agilent strongly
recommends using the parts listed here for all checkout tests.
Agilent also recommends installing new consumable parts
whenever the quality of the installed ones is not known. For
example, installing a new liner and septum ensures that they
will not contribute any contamination to the results.
1 Check the indicators/dates on any gas supply traps.
Replace/recondition expended traps.
2 Install new consumable parts for the inlet and prepare the
correct injector syringe (and needle, as needed).
Table 8 Recommended parts for checkout by inlet type
Recommended part for checkout Part number
Split splitless inlet
Syringe, 10-µL 5181-1267
O-ring 5188-5365
Septum 5183-4757
Liner 5183-4711 (FID)
5062-3587 or 5181-3316
(TCD, uECD, FPD, NPD)
Multimode inlet
Syringe, 10-µL 5181-1267
O-ring 5188-6405
Septum 5183-4757
Liner 5188-6568
Packed column inlet
Syringe, 10-µL 5181-1267
O-ring 5080-8898
Septum 5183-4757
Cool on-column inlet
Septum 5183-4758
Septum nut 19245-80521
Syringe, 5-µL on-column 5182-0836