User guide

62 Advanced User Guide
3 Options
To estimate the actual column length or diameter from an elution
time
1 Set oven ramp 1 to 0.00, then verify that the column is
defined.
2 Perform a run using an unretained compound and record
the elution time.
3 Press [Options], scroll to Calibration and press [Enter].
4 From the calibration list, select the column and press [Enter].
The GC displays the current calibration mode for the
column.
5 To recalibrate or to change calibration mode, press
[Mode/Type] to see the column calibration mode menu.
6 Scroll to Length or Diameter and press [Enter]. The following
choices appear:
Mode
Measured flow
Unretained peak
Calculated length or Calculated diameter
Not calibrated
7 Scroll to Unretained peak and enter the actual elution time
from the run performed above.
8 When you press [Enter], the GC will estimate the column
length or diameter based on the elution time input and will
now use that data for all calculations.
To estimate the actual column length or diameter from the
measured flow rate
1 Set oven ramp 1 to 0.00, then verify that the column is
defined.
2 Set the oven, inlet, and detectors temperatures to 35 °C and
allow them to cool to room temperature.
3 Remove the column from the detector.
CAUTION
When you measure the column flow rate, be sure to convert the
measurement to normal temperature and pressure if your
measurement device does not report data at NTP. If you enter
uncorrected data, the calibration will be wrong.