User guide

56 Advanced User Guide
2 Configuration
Instrument
1 Press [Config]. Scroll to Instrument and press [Enter].
2 Scroll to Serial #. Enter a serial number and press [Enter].
This function can only be done by Agilent service personnel.
3 Scroll to Auto prep run. Press [On/Yes] to enable Auto prep run,
[Off/No] to disable it. See “Pre Run and Prep Run” on
page 171 for details.
4 Scroll to Zero Init Data Files.
Press [On/Yes] to enable it. When it is On, the GC
immediately begins to subtract the current detector
output from all future values. This applies only to digital
output, and is useful when a non- Agilent data system has
problems with baseline data that is non- zero.
Press [Off/No] to disable it. This is appropriate for all
Agilent data systems.
5 Front inlet type: and Back inlet type: are both information
displays. The values are determined by the type of flow
modules installed.
6 The Oven line displays the GC power configuration.
7 Press [Clear] to return to the Config menu or any other
function to end.