User guide

Configuration 2
Advanced User Guide 55
Sample tray (7683 ALS)
This section applies to the 7683 ALS system. To configure the
7693A sampler system, see the 7693A Installation, Operation,
and Maintenance manual.
1 Press [Config][Sample Tray].
2 If the vial gripper is touching vials either too high or too low
for reliable pickup, scroll to Grip offset and press [Mode/Type]
to select:
Up to increase the gripper arm pickup height
Default
Down to decrease the gripper arm pickup height
3 Scroll to Bar Code Reader.
4 Press [On/Yes] or [Off/No] to control the following bar code
setpoints:
Enable 3 of 9—encodes both letters and numbers, plus a few
punctuation marks, and message length can be varied to
suit both the amount of data to be encoded and the space
available
Enable 2 of 5—restricted to numbers but does allow variable
message length
Enable UPC code—restricted to numbers- only with fixed
message length
Enable checksum—verifies that the checksum in the
message matches the checksum calculated from the
message characters, but does not include the checksum
character in the returned message
5 Enter 3 as the BCR Position when the reader is installed in the
front of the tray. Positions 1–19 are available.