46 Advanced User Guide
2 Configuration
To configure a nickel catalyst heater
1 Check that a power source for the Nickel Catalyst heater was
assigned. See “To assign a GC power source to an Aux
thermal zone” on page 45.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1, Thermal
Aux 2, or Thermal Aux 3 depending on where the Nickel
Catalyst heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and select
Nickel catalyst, and press [Enter].
To configure an AED transfer line heater
1 Check that a power source for the AED transfer line heater
was assigned. See “To assign a GC power source to an Aux
thermal zone” on page 45.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1, Thermal
Aux 2, or Thermal Aux 3 depending on where the AED transfer
line heater was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and select
the AED transfer line, and press [Enter].