User guide

Configuration 2
Advanced User Guide 45
Thermal Aux
See “Unconfigured:” on page 26 and “Ignore Ready =” on
page 25.
The auxiliary thermal controllers provide up to three channels
of temperature control. These controllers are labeled Thermal
Aux 1, Thermal Aux 2, and Thermal Aux 3.
To assign a GC power source to an Aux thermal zone
This procedure assigns a the heater power source from heater
plug A1 or A2 to Thermal Aux 1, Thermal Aux 2, or Thermal
Aux 3 temperature control zones.
1 Unlock the GC configuration, press [Options], select Keyboard
& Display and press [Enter]. Scroll down to Hard Configuration
Lock and press [Off/No].
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1, Thermal
Aux 2, or Thermal Aux 3 and press [Enter].
3 With Unconfigured selected, press [Mode/Type], and select:
Install Heater A1 if the heated device is plugged into the
valve box bracket plug labeled A1.
Install Heater A2 if the heated device is plugged into the
valve box bracket plug labeled A2.
4 Press [Enter] after making selection.
5 When prompted by the GC, turn the power off then on again.
To configure an MSD transfer line heater
1 Check that a power source for the MSD heater was assigned.
See “To assign a GC power source to an Aux thermal zone” on
page 45.
2 Press [Config][Aux Temp #] and scroll to Thermal Aux 1, Thermal
Aux 2, or Thermal Aux 3 depending on where the MSD heater
was assigned, and press [Enter].
3 Scroll to Auxiliary type, press [Mode/Type], scroll to and select
the MSD transfer line, and press [Enter].