User guide

Configuration 2
Advanced User Guide 39
The flows to the three detectors are based on the pressure
drops through the capillaries and their resistance to flow. An
Agilent flow calculator provided with the capillary flow splitter
device is used to size the length and diameter of these capillary
sections to obtain the desired split ratios.
Your analytical method can set the flow or pressure for
column # 2, the lowest numbered column in the split. Use the
value obtained from the Agilent flow calculator for this setpoint
in your method.