Setting parameters for auxiliary pressure control
1 Press [Aux EPC #] and scroll to the channel you wish to
control. Press [Enter].
2 Scroll to Initial pressure. Enter a value and press [Enter].
3 If desired, enter a pressure program using the time and rate
functions.
4 Press [On/Yes] to apply Initial pressure and start the program.
About Aux Thermal Zone Control
There can be up to 3 aux thermal controlled zones.
These zones are 3- ramp programmable.
Run time events may be used to schedule specific temperatures
during the run.
Setting parameters for the aux thermal zone control
1 Press [Aux Temp #] and scroll to the zone you wish to control.
Press [Enter].
2 Scroll to Te mp erature. Enter a value and press [Enter].
3 Scroll to Initial time. Enter a value and press [Enter].
4 Scroll to Rate 1. Enter 0 to end the program here or a positive
value to create a temperature program.
About Auxiliary Device Contacts
These contacts are controlled by the external valve drivers. See
“The external valve drivers” on page 342.
About the 24V Auxiliary Device Power Supply
This is controlled by the external valve drivers. See “The
external valve drivers” on page 342.
About Auxiliary Columns
Defines up to 6 columns (includes Col 1 and Col 2).
1 Press [Config][Aux Col #] and enter a column number. Press
[Enter].
2 Define/configure the column. See “Column #” on page 33 for
details.