14 GC Output Signals
About Signals
Signal is the GC output to a data handling device, analog or
digital. It can be a detector output or the output from flow,
temperature, or pressure sensors. Two signal output channels
are provided.
Signal output can be either analog or digital, depending on your
data handling device. Analog output is available at either of two
speeds, suitable to peaks with minimum widths of
0.004 minutes (fast data rate) or 0.01 minutes (normal rate).
Analog output ranges are 0 to 1 V, 0 to 10 V, and 0 to 1 mV.
Digital output rates are set by your Agilent data system, such as
ChemStation or EZChrom.