User guide
Cables 13
Advanced User Guide 365
External event cable
Two passive relay contact closures and two 24- volt control
outputs are provided. Devices connected to the passive contact
closures must be connected to their own power sources.
The pin assignments for this cable are listed in Table 83.
1
2
3
4
5
678
Connector 1
Connector 2
Table 83 External events cable
Connector 1 pin Signal name Maximum rating Connector 2, wire
color
Controlled by valve #
24 volts output
1 24 V output 1 75 mA Yellow 5
2 24 V output 1 75 mA Black 6
3 Ground Red
4 Ground White
Relay contact closures (normally open)
5 Closure 1 48 V AC/DC, 250 mA Orange 7
6 Closure 1 Green 7
7Closure 248 V AC/DC, 250 mA Brown or violet 8
8Closure 2 Blue8