User guide
360 Advanced User Guide
13 Cables
Start (Low True) Request to start run/timetable. Receiver is
any module performing runtime- controlled activities. The 7890
GC requires a pulse duration of at least 500 micro- seconds to
sense a start from an external device.
Start Relay (Contact Closure) A 120 millisecond contact closure
– used as an isolated output to start another device that is not
compatible or connected with APG Remote pin 3.
Stop (Low True) Request to reach system ready state as soon as
possible (for example, stop run, abort or finish, and stop
injection). Receiver is any module performing
runtime- controlled activities. Normally this line is not
connected, if the GC oven program is used to control the
method Stop time.