User guide
Configuration 2
Advanced User Guide 31
To configure the MMI coolant
Press [Config][Front Inlet] or [Config][Back Inlet]. If the inlet has
not been configured previously, a list of available coolants is
displayed. Scroll to the desired coolant and press [Enter].
Cryo type/Cooling type [Mode/Type] displays a list of available
coolants. Scroll to the desired coolant and press [Enter].
Normally, select the coolant type that matches the installed
hardware.
• N2 cryo Select if the N
2
option is installed and you are using
LN
2
or compressed air.
• CO2 cryo Select if the CO
2
option is installed and you are
using LCO
2
or compressed air.
• Compressed air Select if the N
2
or CO
2
option is installed and
you are only using compressed air. If Compressed air is
selected as the Cooling type, air coolant is used to cool the
inlet regardless of the Use cryo temperature setpoint during the
cooling cycle. If the inlet reaches setpoint, the air coolant is
turned off and stays off for the duration of the cooling cycle.
See Cooling the MMI for more information.
If the Cryo type selection is anything other than None, several
other parameters appear.
Cryo [On/Yes] enables cryogenic cooling of the inlet at the
specified Use cryo temperature setpoint, [Off/No] disables cooling.
Use cryo temperature If N2 cryo or CO2 cryo is selected as the
Cryo type, this setpoint determines the temperature below
which cryogenic cooling is used continuously to hold the inlet at
setpoint. Set the Use cryo temperature equal to or higher than the
inlet setpoint to cool the inlet and hold the setpoint until the
inlet temperature program exceeds the Use cryo temperature. If
the Use cryo temperature is less than the inlet setpoint, cryogen
will cool the inlet to the initial setpoint and turn off.
Cryo timeout This parameter is available with N2 cryo and CO2
cryo Cryo types. Use this setting to conserve cryogenic fluid. If
selected, the instrument shuts down the inlet and cryogenic
cooling when no run starts in the number of minutes specified.
The setpoint range is 2 to 120 minutes (default 30 minutes).
Turning cryo timeout off disables this feature. We recommend
cryo timeout enabling to conserve coolant at the end of a
sequence or if automation fails. A Post Sequence method could
also be used.