User guide

Detectors 10
Advanced User Guide 303
3
Set the detector temperature. The temperature must be
greater than 150 °C for the flame to light.
4 Set the hydrogen flow rate, if desired, and press [Off/No].
5 Change the air flow rate, if desired, and press [Off/No].
6 If using a packed column, set the FID makeup gas to 0.0/Off.
7 If using a defined capillary column, set the makeup gas flow
or combined column plus makeup gas flow.
8 Scroll to Flame and press [On/Yes]. This turns on the air and
hydrogen and initiates the ignition sequence. The signal
typically increases to 5 to 20 pA after ignition.
Verify that the flame is lit by holding a cold, shiny surface,
such as a mirror or chrome- plated wrench, over the collector
exit. Steady condensation indicates that the flame is lit.