5
Loosen the hex nut connecting the split vent line to the
interface until you can remove the line. Put the line aside.
You do not need to plug it.
6 Install a blanking nut into the split line port and
finger- tighten the nut. Tighten the nut an additional
1/4- turn using two wrenches in opposition.
7 Place the interface in the heater block. Replace the clamping
plate you removed earlier and tighten the screw until snug.
Do not overtighten. If you removed the transfer line, replace