User guide
26 Advanced User Guide
2 Configuration
[ EPC3 ] = (DET-EPC) (FID) EPC #3 is controlling detector gases
to an FID.
[ EPC6 ] = (AUX_EPC) (PCM) EPC #6 is controlling a
two- channel pressure control module.
FINLET (OK) 68 watts 21.7 This heater is connected to the front
inlet. Status = OK, meaning that it is ready for use. At the time
that the GC was turned on, the heater was drawing 68 watts and
the inlet temperature was 21.7 °C.
[ F-DET ] = (SIGNAL) (FID) The signal board for the front
detector is type FID.
AUX 2 1 watts (No sensor) The AUX 2 heater is either not
installed or not OK.
Unconfigured:
Accessory devices requiring GC power or communication must
be assigned these GC resources before they can be used. To
make this hardware element usable, first “Unlock the GC
Configuration” on page 25 then go to the Unconfigured parameter
and press [Mode/Type] to install it. If the hardware element you
are configuring requires selection of additional parameters, the
GC asks for that selection. If no parameters are required, press
[Enter] at the GC prompt to install that element. You are
required to power the GC off and then power the GC on to
complete this configuration.
After restarting the GC, a message reminding you of this change
and its effect on the default method is displayed. If needed,
change your methods to accommodate the new hardware.