User guide

254 Advanced User Guide
8 Inlets
After timing a trial set of 10 injections, the total time for the
multiple injection set was measured to be approximately
1.3 minutes. The following timing changes were made:
The result is shown in the next figure. Note the difference in the
vertical scale (5000 versus 500).
Table 50 Modifications
Parameter Increased from To
Inlet Init time 0.3 minutes 1.6 minutes
Vent end time 0.2 minutes 1.5 minutes
Purge time 2.0 minutes 3.0 minutes
Oven Init time 2.5 minutes 3.0 minutes
C20