User guide

Inlets 8
Advanced User Guide 249
Enter a Vent pressure of 0 psi (0 kPa) and Vent end time of
0.1 minutes.
d Make an injection using the empty vial. Use a stopwatch
or your GC's timer feature to time how long the needle is
in the inlet.
8 Enter revised Solvent vent mode parameters.
Set the method's Vent end time to be about 0.05 min longer
than the time the needle spends in the inlet.
Set the inlet temperature Initial time to be about 0.05 min
longer than the Vent pressure time.
Program the inlet to ramp quickly to the injection
temperature. Make sure the ramp starts after the Vent end
time.
Set the Purge Flow to 30 mL/min. Set the purge time to be
the vent pressure Vent end time + 1 minute.
9 Make a 5 uL injection of your standard. You should see 5
times the response.
If the response of all analytes is too low:
The dispense speed is too fast. Liquid was injected into
the inlet and pushed out the vent.
The vent time is too long. The inlet started to heat while
the vent was open.
If the response of early eluters is too low:
The inlet temperature is too high.
The Vent flow is too high.
If the response of late eluters is too low:
The purge time is too short.
The final inlet temperature is too low.
10 If you need more injection volume and the 5 uL worked to
give 5 times the response, change the syringe to a larger one,
for example, 50 uL.
11 Set up the data system to perform a 25 uL injection.
12 Configure the syringe. Make sure the plunger speeds on the
injector are still set properly.
13 Recheck the injector timings. See the step above.
14 Set a new vent time and initial inlet temperature time based
on the new time the needle spends in the syringe.
15 Make a 25 uL injection of your standard. Again, you should
see 5 times the response. If not, see step 13.