User guide

244 Advanced User Guide
8 Inlets
When is Start Run?
Both the inlet and oven temperature programs begin at Start
Run. All times—such as Purge time—are measured from Start
Run. When does Start Run occur?
If the sample is injected manually, Start Run occurs when the
user presses the [Start] key.
If a single injection per run is made using an autosampler,
Start Run occurs when the syringe carrier moves down to
make the injection.
If multiple injections per run are made using an autosampler,
Start Run occurs when the syringe carrier moves down to
make the first injection of the set. There are no Start Run
signals for the rest of the injections in the set.
These additional injections take time. The inlet and oven
temperature programs, mainly the Initial time values, must be
adjusted to allow for this. So must the various time values that