Cold split introduction
For cold split sample introduction, use an initial inlet
temperature below the normal boiling point of the solvent. If the
liner volume is enough to hold all the vaporized solvent, start
the first inlet temperature ramp at 0.1 minutes with a high
heating rate (500 °C/min or higher). The final temperature
should be high enough to volatilize the heaviest analytes from
the liner and should be held for at least 5 minutes. A final
temperature of 350 °C for 5 minutes has proven sufficient to
quantitatively transfer C
44
.
For larger injection volumes or to eliminate the solvent, hold the
initial temperature long enough to vent the solvent through the
split vent and then begin the first ramp. Use a fast rate for
thermally stable analytes. Slower rates may help minimize
thermal degradation in the inlet.