Setting parameters for the inlet in solvent vent mode
Set or configure the following parameters in the data system's
7890A GC method editor.
Syringe size — Verify the syringe size is configured correctly.
The configured syringe size changes the available choices for
injection volume.
Injection volume — Select the injection volume, then enter a
number of injections. The total injection volume will be
displayed.
Multiple Injection Delay — A pause time, in seconds, between
injections. This is added to the minimum hardware cycle time.
Preinjection washes and pumps are performed only before the
first injection of a multiple injection set.
Postinjection washes are performed only after the last injection
in a multiple injection set.
An example
These values were used for a sample with a broad range of
boiling points.
Table 38 General parameters
Name Value
Sample C
10
to C
44
hydrocarbons in hexane
Mode Solvent vent
MMI liner Glass wool packed
Injection volume One 10.0 μL injection (25 μL syringe)
Injection speed Fast
Column 30 m x 320 μm x 0.25 μm -5, part
number 19091J-413
Column flow 4 mL/min constant flow
Table 39 Inlet parameters
Name Value Name Value
Initial temp 40 °C Rate 2 (off)
Initial time 0.3 min Pressure 15.6 psig
Rate 1 720 °C/min Vent pressure 0.0 psig