User guide

212 Advanced User Guide
8 Inlets
Start with the inlet temperature cold, near but slightly
below the solvent boiling point. For example, if using
methylene chloride (boiling point 39 °C), start with a
temperature of 30–39 °C.
Use Splitless mode.
Ramp to the normal split/splitless inlet temperature
5 Note the response achieved.
6 Next, change the inlet mode to Solvent Vent.
7 Check the injector timings.
a Install an empty sample vial in the injector turret or tray.
b Input the injection rate (Draw, Dispense, and Inject rates)
for the injector and increase the injection volume to 5 uL.
c Enter draft Solvent Vent parameters:
Set a Vent Flow of 100 mL/min as a starting point.
Keep the inlet isothermal for now
Enter a Vent Pressure of 0 psi (0 kPa) until 0.1 min.
Make an injection using the empty vial. Use a
stopwatch or your GC's timer feature to time how long
the needle is in the inlet.
8 Enter revised Solvent Vent mode parameters.
Set the method's Vent Time to be about 0.05 min longer
than the time the needle spends in the inlet.
Set the inlet temperature initial Hold Time to be about
0.05 min longer than the vent pressure until time.
Program the inlet to ramp quickly to the injection
temperature. Make sure the ramp starts after the vent
pressure until time.
Set the Purge Flow to Split Vent to 30 mL/min. Set the
purge time to be the vent pressure until time + 1 minute.
9 Make a 5 uL injection of your standard. You should see 5
times the response.
If the response of all analytes is too low:
The dispense speed is too fast. Liquid was injected into
the inlet and pushed out the vent.
The vent time is too long. The inlet started to heat
while the vent was open.
If the response of early eluters is too low:
The inlet temperature is too high.