You can do column pressure and flow programming when in the
pressure pulse mode. However, the pressure pulse will take
precedence over the column pressure or flow ramp.
Split/Splitless inlet split mode minimum operating pressures
The minimum recommended inlet total flow is 20 mL/minute.
When the split/splitless inlet is operated in Split mode, there
will be a minimum pressure at which the inlet can operate.
Typically, low inlet pressures may be required for shorter, wide
bore columns. The minimum pressure is a function of carrier
gas type, total inlet flow, liner design, and possible
contamination of the split vent tube or trap.
A wide bore column requires a much lower inlet pressure than a
typical capillary column to maintain a given flow. Setting the
split ratio (total flow) too high when using a wide bore column
can create an unstable control relationship between the
pressure and flow control loops.