User guide

Advanced User Guide 17
Clock Time Programming
Clock time programming allows certain setpoints to change
automatically at a specified time during a 24- hour day. Thus, an
event programmed to occur at 14:35 hours will occur at 2:35 in
the afternoon. A running analysis or sequence has precedence
over any clock table events occurring during this time. Such
events are not executed.
Possible clock time events include:
Valve control
Method and sequence loading
Starting sequences
Initiating blank and prep runs
Column compensation changes
Adjustments of the detector offset
Using clock time events
The Clock Table function allows you to program events to occur
during a day based on the 24- hour clock. Clock table events that
would occur during a run or sequence are ignored.
For example, the clock table could be used to make a blank run
before you even get to work in the morning.
Programming clock time events
1 Press [Clock Table].
2 Press [Mode/Type] to see the available clock time events.
3 Scroll to the parameter to be programmed.
4 Edit Time: and the setpoints for this event.
5 Press [Mode/Type] to add another event. Press [Status] to
terminate entries.
When the clock event is executed, a confirming message
appears.
Adding events to the clock table
1 Press [Clock Table].
2 Press [Mode/Type]. When entries are added, they are
automatically ordered chronologically.
3 Select the event type.
4 Set appropriate parameters.