6 Checking for Leaks
To Correct Leaks in the PTV Inlet
If the inlet fails a pressure decay test, check the following:
Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
If using a septum head, check the tightness of the septum
nut. See To change the septum on the PTV inlet.
If using a septum head, check the septum. Replace if old or
damaged.
Check that the septumless- or septum- head assembly is
installed tightly.
Replace the liner and Graphpak 3D ferrule. See To change
the liner on the PTV inlet.
Check the Graphpak adapter seal against the inlet body.
Replace the silver seal and reinstall the adapter if needed.
See To replace the inlet adapter for the PTV inlet.
If using the septumless head, check for leaks around the
guide cap. Replace the Teflon ferrule in the guide cap. See To
replace the Teflon ferrule on a PTV inlet.
Check for leaks at the split vent trap. Tighten as necessary.
Replace the split vent trap filter and O- rings. See To replace
the filter in the PTV inlet split vent line.
If these items not resolve the problem, contact Agilent for
service.