User guide

Checking for Leaks 6
Advanced User Guide 137
To Correct Leaks in the Packed Column Inlet
If the inlet fails a pressure decay test, check the following:
Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
Check the tightness of the septum nut. See To change the
septum on the purged packed inlet.
Check the septum. Replace if old or damaged.
Check that the top insert weldment is installed tightly.
Replace the O- ring. See To change the O- ring on the purged
packed inlet. Also check the glass insert. See To change the
glass insert on a PP inlet.
Replace the ferrule seal on the adapter.
If these items do not resolve the problem, contact Agilent for
service.