User guide

132 Advanced User Guide
6 Checking for Leaks
To Correct Leaks in the Multimode Inlet
If the inlet fails a pressure decay test, check the following:
Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
Check the tightness of the septum nut. See To change the
septum on the Multimode inlet.
Check the septum. Replace if old or damaged.
Check the insert assembly installation.
Check the liner and liner O- ring. See To change the liner and
O- ring on the Multimode inlet.
If these items do not resolve the problem, contact Agilent for
service.