To program an ALS subsequence
1 See step 1 through step 3 of “To program a sequence” on
page 111.
2 Press [Mode/Type] and select the injector type.
3 Enter injector sequence parameters (if using both injectors,
there will be two sets of parameters):
• Number of Injections/vial—the number of repeat runs from
each vial. Enter 0 if no samples are to be injected.
• Samples—the range (first–last) of sample vials to be
analyzed.
4 Proceed with step 5 of “To program a sequence” on page 111.
To program a valve subsequence
1 See step 1 through step 3 of “To program a sequence” on
page 111.
2 Press [Mode/Type] and select Valve.
3 Enter the valve sequence parameters (the first three appear
only if a multiposition valve is configured):
• #inj/position—number of injections at each position (0–99)
• Position rng—first–last valve positions to sample (1–32)
• Times thru range—number of times to repeat the range
(1–99)
• # injections—number of injections for each sample
4 Proceed with step 5 of “To program a sequence” on page 111.
To program post sequence events
1 See step 1 through step 4 of “To program a sequence” on
page 111.
2 Scroll to the Method # line of Post Sequence and enter a
method number. Use 1 to 9 for the stored methods, or 0 if
there is no method to be loaded.
3 Press [On/Yes] at Repeat sequence to keep repeating the sequence
(useful for valve sequences). Otherwise, press [Off/No] to halt the
sequence when all subsequences are finished.