User guide

Methods and Sequences 5
Advanced User Guide 109
Enable 3 of 9—encodes both letters and numbers, plus a few
punctuation marks, and message length can be varied to
suit both the amount of data to be encoded and the space
available
Enable 2 of 5—restricted to numbers but does allow variable
message length
Enable UPC code—restricted to numbers- only with fixed
message length
Enable checksum—verifies that the checksum in the
message matches the checksum calculated from the
message characters, but does not include the checksum
character in the returned message
4 Enter 3 as the BCR Position when the reader is installed in the
front of the tray. Positions 1–19 are available.
To save a method
1 Press [Method] and scroll to the desired method number.
2 Press [Store] and [On/Yes] to store the new method using the
chosen number. Alternatively, press [Off/No] to return to the
stored methods list without saving the method.
A message is displayed if a method with the number you
selected already exists.
Press [On/Yes] to replace the existing method or [Off/No]
to return to the stored methods list without saving the
method.
To load a stored method
Press [Load][Method]. Supply the method number and press
[Enter]. The specified method will replace the current active
method.
Method mismatch
This section applies only to a standalone (not connected to a
data system) GC. When a data system, such as a ChemStation or
EZChrom Elite, controls the GC, methods are stored in the data
system and can be edited there. See your data system
documentation for more information.