Ignite the FPD flame, if not lit.
8 Display the signal output and monitor. This output typically
runs between 40 and 55 but can be as high as 70. Wait for the
output to level off. This takes approximately 1 hour.
If the baseline output is too high:
Check column installation. If installed high, the stationery
phase burns out and increases measured output.
Check for leaks.
Bake out the detector and column at 250 °C.
Wrong flows set for installed filter
Sample washes 2
Sample pumps 6
Sample wash volume 8
Injection volume 1 µL
Syringe size 10 µL
Solvent A pre washes 2
Solvent A post washes 2
Solvent A wash volume 8
Solvent B pre washes 0
Solvent B post washes 0
Solvent B wash volume 0
Injection mode (7693A) Normal
Airgap Volume (7693A) 0.20
Viscosity delay 0
Inject Dispense Speed (7693A) 6000
Plunger speed (7683) Fast, for all inlets except COC.
PreInjection dwell 0
PostInjection dwell 0
Manual injection
Injection volume 1 µL
Data System
Data rate 5 Hz
Table 17 FPD Phosphorus Checkout Conditions (continued)