User guide
Multimode Source for 6100 Series Single Quad LC/MS User Guide 9
Basic Operation and Maintenance 1
To create a method for positive/negative mixed mode operation
To create a method for positive/negative mixed mode operation
1 Open the MSD Spray Chamber dialog box by clicking the Instrument > Set
Up MSD Signals in the Method and Run Control view.
2 Select MM-ES_APCI from the Method Spray Chamber drop-down list.
3 Check that Installed Spray Chamber is also set to MM-ES+APCI.
4 Make any other changes that are necessary for your method.
5 Click the OK button.
6 Open the Set up MSD Signals dialog box by clicking the Instrument > MSD
Spray chamber in the Method and Run Control view.
7 Modify the settings so that Signal 1 has Positive polarity and Signal 2 has
Negative polarity as shown in Figure 1.
8 Make any other changes that are necessary for your method.
9 Click the OK button.
Rapid polarity switching is a very useful technique but it requires time for the
ion chemistry to be established and the optics path to refill with ions. The gas
density plays a role in the speed of refilling the ion path. The gas density is
affected by source temperature. For a method running positive/negative
switching, it is advised to use a lower vaporizer temperature (150 to 200°C)
and a lower Vcap (approximately 1000 V). These will greatly affect the quality
of the results in positive/negative switching experiments.