G2422A MM-APCI Calibration Ions
G2422A MM-APCI Calibration Ions
NOTE
Fragment ion signals are progressively stronger at higher m/z by increasing CID energy.
Positive Ion (m/z) Empirical Formula
Purine+H 121.051421 C5.H5.N4
HP-0321+H 322.048699 C6.H19.O6.N3.P3
HP-0622+H 622.029499 C12.H19.O6.N3.P3.F12
HP-0921+H 922.010300 C18.H19.O6.N3.P3.F24
fragment 1307.969049 C25.H17.O6.N3.P3.F40
HP-1521+H 1521.971900 C30.H19.O6.N3.P3.F48
fragment 1807.937049 C35.H17.O6.N3.P3.F60
HP-2121+H 2121.933500 C42.H19.O6.N3.P3.F72
Negative Ion (m/z) Empirical Formula
Purine-H 119.035771 C5.H3.N4
fragment 556.001426 C10.H15.O6.N3.P3.F10
fragment 805.985476 C15.H15.O6.N3.P3.F20
fragment 1305.953400 C25.H15.O6.N3.P3.F40
fragment 1805.921399 C35.H15.O6.N3.P3.F60