Multimode Source for 6100 Series Single Quad LC/MS User Guide 61
Reference 3
Guidelines for positive/negative polarity switching in mixed mode
Guidelines for positive/negative polarity switching in mixed mode
The following settings are a good starting point when you are doing
positive/negative polarity switching in MM-ES+APCI mode.
Drying gas flow Always set to 5 L/min (IR does the drying).
Nebulizer pressure Start at 40 psi. You will use the nebulizer pressure and the
corona current to balance ESI and APCI. A higher nebulizer pressure results
in more ESI. A lower nebulizer pressure results in more APCI.
Drying gas temperature Set to 250°C. Remember you want the vaporizer to
control the drying and to be able to maintain the set point.
Vaporizer Start at 150°C. A higher temperature usually works better for
MM-APCI. MM-ESI is usually unaffected except for sodium adduction. The
vaporizer will compensate for flow rate or solvent composition.
Capillary Voltage Set to 1000 V. The lower voltage works better for fast
switching. You are balancing ESI and APCI response.
Corona current Set to 1 µA for positive or negative polarity. The corona
voltage will be approximately 3500 at 2 µA. More corona current will reduce
the ESI signal and increase the APCI signal.
Charging Voltage Set the charging voltage to 2000 V.
NOTE
Click Instrument > MSD Spray Chamber to set the Method Spray Chamber to
MM-ES+APCI before you edit the Set Up MSD Signals dialog box.
NOTE
Refer to “To create a method for positive/negative mixed mode operation” on page 9 for
additional information.